MACS Matchmaker
Plan a MACS Matchmaker measurement: choose what to immobilize, set up assay phases, build injections in the Injection Manager, and configure flow and buffer conditions. Advanced topics — Rmax estimation, ligand density, pickup volume, and regeneration — are in the expandable section at the bottom.
Choosing Ligand and Analyte
The decision of which molecule to immobilize on the mologram (ligand) and which to flow in solution (analyte) affects data quality and interpretation. Consider the following guidelines:
- Immobilize the smaller, more stable molecule. A smaller ligand leaves more room for the larger analyte to bind without steric interference. Additionally, larger analytes produce a stronger signal per binding event (more pg/mm² per molecule), improving the signal-to-noise ratio.
- Flow bivalent molecules as analyte with caution. Bivalent analytes (e.g., full IgG antibodies) can bind two adjacent ligands simultaneously, introducing avidity effects that complicate kinetic analysis. If avidity is a concern, consider using monovalent fragments (Fab) as the analyte, or reduce ligand density to minimize the probability of bivalent binding.
- Preserve ligand activity. The immobilized molecule must remain functional after coupling. For click chemistry chips, covalent attachment via TCO-Tz/MeTz may affect protein folding if the label is placed near the binding site. For DDI chips, the DNA conjugation strategy is generally milder and more predictable.
- Practical constraints. Immobilize the molecule that is available in smaller quantity (less material is needed for surface immobilization than for flowing multiple concentrations). Flow the molecule that is easy to prepare in a dilution series.
Assay Phases
Each assay type starts from a phase template. Use the guidance below for phase-specific design choices; for all flow rates and injection volumes, see Recommended Flow Parameters.
Baseline
Allows the system to stabilize before the first relevant injection. Run the running buffer until the baseline is flat.
Association & Immobilization
- Frontfilling and backfilling can be configured with the Immobilization template when appropriate (see Immobilizations).
- In the guided experiment wizard, choosing the [oligo | oligo] sensor architecture adds an empty ligand row to the Samples step with its placement set to Backfilling.
- For titration experiments (LOD, SCK, MCK, …) the sample/target must be defined as the Analyte.
- To tune ligand surface density on DDI chips, see Ligand Density vs. Mass Transport Trade-Off.
Kinetics Injections
- Use a kinetics phase template when it matches your experiment.
- Typical durations: 2 min association, 2 min dissociation.
Scouting
Use Scouting to confirm that your assay works before committing instrument time to a full experiment: that the ligands immobilize, that backfilling behaves as expected, and that the analyte binds. It also gives a quick yes/no answer on whether a ligand binds its analyte. Backfilling and analyte injection are both optional. When multiple analytes are configured, every Scouting cycle injects them in table order, using the same association and dissociation sequence for each.
Ligands are grouped in table order. A new Scouting cycle starts whenever any placement, including backfilling, appears again. When Inject as Pool is enabled, the non-backfilling ligands in each cycle are combined into one injection.
Scouting cycles omit leading run-buffer commands from immobilization, association, and regeneration phases when the preceding phase already ends with run buffer. Repeated Immobilization, Scouting, and Regeneration phases receive numbered names so that every phase name is unique.
Limit of Detection (LOD)
Experimental setup for LOD measurements. For analysis parameters and the LOD calculation method, see LOD Evaluation.
- Use an LOD-oriented phase template if available.
- Blanks: > 5 times.
- Typical durations: 2 min baseline, 5 min association.
- Flush sample after association: 200 µL flush at 200 µL/min.
- Use the same flow rate for association and dissociation to minimize noise impact.
Concentration Determination
- Use a concentration-determination phase template if available.
- Standards: > 5 different concentrations, duplicates recommended.
Experiment Planning
Titration Series
When planning an experiment, injections are organized into phases. One such phase type is the 'Titration Series'. To use the Titration Series, follow the steps outlined below:
- Create a new sample: This sample will serve as the base for all diluted samples, which are auto-generated by the Titration Series dialog. The color of the base sample will be used to distinguish the diluted samples.
- Open the Titration Series dialog:In the 'Injections' section, click on the round plus button to add a new phase and select 'Titration Series'.
- Fill in the dialog:Specify the highest concentration, the dilution factor, the number of dilution steps, and the placement of blanks. It is possible to repeat the titration series multiple times by setting the repetitions field. A preview of the injection groups with their respective concentrations will appear in the 'Titration Step' section.
- Create the step template:Add commands to the step template. These commands will be replicated in each step with the adjusted concentrations. You can choose from command types 'Run Buffer', 'Inject Titration Sample', and 'Inject Sample'. Use 'Inject Titration Sample' for injecting diluted samples, and 'Inject Sample' for regeneration steps where the same solution is injected at each step.
- Press 'Create New Phase': This action creates the new diluted samples and one injection group for each dilution step. Manual adjustments to the injection groups, such as adding more blanks or changing the order of injections, are possible.
- Press 'Auto Place Vials':After making adjustments, press the 'Auto Place Vials' button in the 'Well Plates' section to place the vials on the Well Plate. Use drag and drop to move the vials to the desired positions.
For guidance on choosing the highest concentration, dilution factor, and number of steps — particularly when is unknown — see Choosing Concentrations.
Single Cycle Kinetics
What it is: Single Cycle Kinetics (SCK) is a series of analyte injections with increasing concentrations in one continuous kinetic run. It is a special case of the Titration Series and can use a single expected value to plan the concentration sweep.
How to design it: choose the analyte sample, open the Single Cycle Kinetics dialog, enter the expected or manually set the concentration range, then verify association, dissociation, and regeneration timing before creating the phase:
- Create a new sample: This sample will serve as the base for all diluted samples, which are auto-generated by the single cycle kinetics dialog. The color of the base sample will be used to distinguish the diluted samples.
- Open the Single Cycle Kinetics dialog:In the 'Injections' section, click on the round plus button to add a new phase and select 'Single Cycle Kinetics'.
- Fill in the dialog: Specify the expected value and press the arrow next to it to fill in the concentration sweep boundaries. The dialog will calculate the suggested concentrations for the injections based on the given dilution factor.
Alternatively, the concentration range can also be set manually by filling the Start/End fields. - Verify interaction times You should now see the step template with the association and dissociation injection for each concentration. Adjust the duration of these steps and duration of the last regeneration step according to your expected kinetics.
- Generate the PhaseAs with the titration series you can now generate the phase by pressing 'Create New Phase' and afterwards auto place the vials with 'Auto Place Vials'.
Choosing Concentrations
If the is known, enter it in the dialog and the software will suggest an appropriate concentration range. If the is unknown, use the following guidelines:
- Start with a wide range, e.g., 0.1 nM to 1 µM, using a 3-fold dilution series (5–7 concentrations).
- Use literature values or prior experiments to estimate the range.
- Aim to include concentrations both well below and well above the expected to capture the full binding curve from near-zero to near-saturation.
- After a first scouting experiment, narrow the range around the observed for more precise fitting.
Why Concentrations Must Span KD
The binding isotherm is most sensitive to in the 20–80% saturation window (approximately 0.1× to 10× ). If all concentrations are too high (well above ), the sensor reaches saturation at every step and the curves become indistinguishable — the fit cannot resolve or accurately. If all concentrations are too low (well below ), the responses are small and dominated by noise, making the fit unreliable. A well-designed concentration series includes points that produce clearly sub-saturating responses, near- responses, and near-saturating responses.
Multi Cycle Kinetics
What it is: Multi Cycle Kinetics (MCK) repeats analyte injections across separate association and dissociation cycles. Use MCK when regeneration between concentrations is required or when each concentration should start from a fresh baseline.
How to design it: define the analyte sample, open the Multi Cycle Kinetics dialog, configure the concentration sweep, then review the generated baseline, association, dissociation, and regeneration steps:
- Create a new sample: Define the analyte sample that will be used across the concentration series.
- Open the Multi Cycle Kinetics dialog:In the 'Injections' section, click the round plus button and select 'Multi Cycle Kinetics'.
- Configure the concentration sweep: The dialog behaves like a titration series and creates one measurement cycle per concentration. Review the suggested concentrations and repetitions before creating the phase.
- Adjust the step template: Include the required baseline, immobilization, association, dissociation, and regeneration commands. The standard template starts with a 2 min baseline at 300 µL/min, injects ligand for 5 min at 20 µL/min, adds a 3 min baseline at 300 µL/min, injects the titration sample for 2 min at the flow-chamber-dependent association flow rate (30 µL/min for the 6-channel chamber, 60 µL/min for the 3- and 4-channel chambers, and 100 µL/min for the single-channel chamber). The wizard enforces a minimum association flow rate of 60% of that default, rounded up to the next 10 µL/min. The command uses 30 µL extra pickup, dissociates for 2 min at 200 µL/min, and regenerates for 0.5 min at 400 µL/min.
- Generate and review the phase: Create the phase, verify the resulting injection groups, and auto place vials once the layout is finalized.
The same concentration-range logic applies to MCK: span with sub-saturating, near-, and near-saturating points (see Choosing Concentrations).
Injection Commands
Inject and Incubate
When selecting the pickup volume for the inject and incubate command, consider the effects of diffusion from the autosampler loop. As illustrated in the figure below, a larger volume reduces the impact of diffusion. Both the incubation duration and fluid viscosity influence this process. Given a sensor volume of 3 µL in the single flow chamber, a default value of 30 µL is recommended.

Inject with Alternate Flow
If the sample volume is limited and/or a high flow rate is required, the alternate flow command can be used. This pushes and pulls the sample over the sensor for a specified time. The pickup volume is the amount of sample that is delivered to the sensor. The displacement volume defines how much the sample is pushed forth and back. Both parameters can be set in the command dialog. Large displacement volumes lead to loss of the sample after a few minutes due to the dilution and diffusion effects. A displacement volume of less than 10 µL is recommended to avoid excessive diffusion. For an incubation of more than 5 minutes, a displacement volume of 1-2 µL is recommended.

Other Commands
- Inject Sample: injects a sample. A vial is automatically placed at the next free location and the required volume is calculated when Auto Place Vialsis pressed. The required volume equals the injection's pickup volume, which includes the configured additional pickup volume and, for partial-loop pickup volumes from 227 to 250 µL, a non-editable 30 µL loop flush the autosampler needs in partial loop mode. Vial dead volume is not accounted for.
- Run Buffer: runs the running buffer for the specified time and flow rate.
- Inject from Vial: injects a user-specified volume from a user-specified vial. The vial must be filled with sample before the injection — no auto-placement is performed.
Stop Condition
Each command includes an optional Stop Condition that automatically terminates the command based on specific criteria. There are three selectable stop conditions:
- None: Default setting. The command runs until completion unless manually edited or fast-forwarded.
- Plateau: Monitors the median coherent mass density of active signal molograms. The system records 10 initial data points, then evaluates whether the last 4 points deviate less than 1% from their median. When this criterion is met, the command fast-forwards.
- Level Reached: Terminates when the median coherent mass density reaches a configured target. Three sub-options select how the target is interpreted:
- ↑ Rise (association): the value entered is a delta (Δ); the target is starting signal + Δ. The command stops when the last 2 points (in pg/mm²) are all at or above the target.
- ↓ Fall (dissociation): the value entered is a delta (Δ); the target is starting signal − Δ. The command stops when the last 2 points are all at or below the target.
- ≷ Absolute value (auto-detect direction): the value entered is treated as an absolute target in pg/mm² rather than a delta. The direction is inferred from the signal at the start of the command: if the initial 4-sample window sits below the target the command stops on an upcross; if it sits above, the command stops on a downcross.
Examples (Δ-based): if the baseline is 40 pg/mm² and you enter a magnitude of 25 with 'Rise', the command stops once the median reaches 65 pg/mm². With 'Fall' it would stop at 15 pg/mm².
Example (absolute): with an 'Absolute value' target of 5 pg/mm² and an initial signal of ~40 pg/mm² (higher than the target), the command stops when the signal falls to 5 pg/mm² — useful as a safety floor that prevents, e.g., zero-crossing.

Autosampler Injection Modes
The MACS Sampler supports three injection modes. The software automatically selects the best mode based on the programmed injection volume:
| Mode | Volume range | Reproducibility | Sample loss |
|---|---|---|---|
| µL Pickup | Up to 227 µL | < 1% RSD | None |
| Partial loop | 227–250 µL | < 0.5% RSD | 30 µL (flush volume) |
| Full loop | 500 µL (loop volume) | < 0.3% RSD | 230 µL |
RSD = Relative Standard Deviation, a measure of injection reproducibility (lower is better).
Full Loop Injections
The sample loop is completely filled (quantitatively) with sample. This type of injection results in extremely good reproducibility.
- Step 1: At the beginning, the injection valve is in INJECT position. The sample needle with air needle has entered the well or vial. Headspace pressure, applied through the air needle, ensures that no air or vapor bubbles are formed during sample aspiration.

- Step 2: The syringe dispenser aspirates the flush volume from the sample well/vial to fill the sample line with sample and remove wash solvent.

- Step 3: The injection valve is switched to the LOAD position, placing a distinct sample plug at the inlet of the sample loop.

- Step 4: The sample loop is quantitatively filled by transporting a number of times the loop volume through the loop, depending on the volume of the loop:
- 3 x loop volume for loops ≤ 100 µL
- 2 x loop volume for loops between 100 µL and 500 µL
- 1.5 x loop volume for loops > 500 µL

- Step 5: The injection valve switches to the INJECT position. The sample loop is not part of the HPLC mobile phase flow path: sample is transported to the column. The analysis starts.

A wash routine is performed after each injection.
Partial Loop Fill Injections
The switching sequence for a partial loop fill injection is:
- Step 1: At the beginning, the injection valve is in the INJECT position. The sample needle with air needle has entered the vial/well. Headspace pressure (optional), applied through the outer air needle, ensures that no air or vapor bubbles are formed during sample aspiration.

- Step 2: The syringe dispenser aspirates the flush volume from the sample vial to fill the sample line with sample and remove wash solvent.

- Step 3: The injection valve switches to LOAD, ensuring a distinct sample plug at the beginning of the sample loop.

- Step 4: The programmed injection volume is now aspirated into the sample loop.

- Step 5: The injection valve switches to INJECT. The sample loop is now part of the (U)HPLC mobile phase flow path: sample is transported to the column. The analysis starts.

If an injection from the same vial and no wash routine is programmed, the next injection sequence will start with a flush of 50% of the programmed flush volume. Otherwise, it will start with a flush of the programmed flush volume. If the withdrawal of sample for the next injection exceeds the total volume of the sample buffer tubing, the buffer tubing is rinsed before the next injection. The next injection will start with the programmed flush.
µL Pickup Injections
The switching sequence for µL pickup injections is:
- Step 1: At the beginning, the injection valve is in INJECT position. The sample needle has entered the transport position.
- Step 2: The transport reservoir is filled with wash solvent the programmed amount of times the syringe volume (after a wash or after emptying of buffer tubing). The injector valve remains in INJECT position during fill transport. Please note that the transport solvent needs to be compatible with eluent.
- Step 3: For the first injection, the syringe dispenser aspirates a transport plug from the transport position to fill the sample line with transport liquid and remove wash solvents.
- Step 4: The needle moves from the transport position to the sample vial. The injection valve switches to LOAD position.

- Step 5: The programmed injection volume is aspirated from the sample vial.

- Step 6: The sample needle moves back to the transport position. A second transport plug is aspirated. The sample is quantitatively transported into the loop.

- Step 7: The injection valve switches to INJECT. The sample loop is now part of the HPLC mobile phase flow path: sample is transported to the column. The analysis timer starts.


The sequence is repeated for each injection.
Samples and Buffer Recommendations
- Focal molography is more tolerant of sample-to-running-buffer mismatch than SPR, so exact matching is not always required.
- For quantitative kinetics and whenever bulk effects are visible, still minimize mismatch between running buffer and sample buffer.
- Suitable buffers: PBS, HEPES, or Tris-based systems. All buffers should be supplemented with surfactant molecules, such as Tween 20, at a recommended concentration of 0.05% v/v.
- Buffers and samples can also contain pre-blocking molecules, such as BSA, casein, or MoloBlock.
- For complex media samples, we recommend centrifuging samples and using only supernatant to prevent clogging of the tubing with aggregated sample.
- Avoid particulate matter in samples to prevent clogging of the fluidic system.
Recommended Flow Parameters
The MACS Matchmaker is compatible with aqueous solutions from pH 1–13. For flow chamber layouts and mologram counts, see Chip Assembly.
Default values: use the preferred-flow column as the starting point for new experiments, then adjust if sample volume, kinetics, mass transport, or regeneration strength require it.
| Phase | Minimum flow | Preferred flow | Notes |
|---|---|---|---|
| Baseline | 200 µL/min | 300 µL/min | 3–5 min to stabilize before the first injection. |
| Immobilization (DDI) | 20 µL/min | 20 µL/min | 10 min at 100–200 nM ligand concentration. |
| Immobilization (TCO–MeTz / TCO–Tz) | 6 µL/min | 6 µL/min | ~15 min, or until plateau is reached. |
| Association (kinetics) | 30 µL/min | 50 µL/min or higher | Use higher flow to reduce mass transport limitations. |
| Dissociation (kinetics) | 100 µL/min | 200 µL/min | Higher flow when stronger washout is needed. |
| LOD association | 20 µL/min | Match association & dissociation | Keep association and dissociation flows equal to minimize noise impact. |
| Regeneration | 400 µL/min | 400 µL/min | Typically 30 s, followed by 3 min running buffer at 300 µL/min. |
Injection volume:use > 90 µL (avoid < 50 µL) to prevent sample dilution from tail diffusion.
Effect of Flow Rate on Association Kinetics
The overlay below shows the same association phase at different flow rates. If the sensogram shape changes when you increase the flow rate, mass transport is likely limiting your measurement — continue increasing until the curves overlay.

Measuring in complex media
Complex media — serum, plasma, cell lysate, conditioned media, crude supernatant — carry high-abundance background proteins in the µM–mM range. Focal molography is built for these samples: coherent non-specific binding is rejected by affinity-matched backfilling, and the ridge–groove self-reference cancels the bulk and thermal terms that swamp refractometric sensors. As a result you can usually measure in undiluted matrix, without the dilution, washing, and proofreading steps a refractometric assay needs. The guidelines below turn that capability into reliable numbers.
1. Start with a backfilled chip
Use an affinity-matched (backfilled) sensor: [oligo | oligo] for regenerable DDI assays or [MeTz | Tz] for covalent click chemistry. A passivated [… | PEG] chip is only appropriate in buffer. Pick the product from the Chip Selection Guide; for custom ligands, design the backfill to the criteria under Choosing a backfill probe.
2. Prepare the sample
- Clarify:centrifuge and use only the supernatant; filter through 0.2 µm where the sample allows. Particles must stay below 50 µm to avoid clogging the fluidics.
- Keep it cold:hold samples at 4 °C in the MACS Sampler to limit degradation and protease activity during long titrations.
- Mind viscosity: plasma and serum are more viscous than buffer, which thickens the diffusion boundary layer and can push the assay toward mass-transport limitation — account for it when interpreting on-rates.
3. Buffer and blocking
Run a buffer that matches the sample’s ionic strength and pH, with 0.05% Tween 20 to suppress non-specific adsorption and air bubbles. For serum or plasma, add a blocker such as 0.1% BSA (or casein / MoloBlock) to the buffer and, where appropriate, the sample. See Samples and Buffer Recommendations.
4. Run the right controls
- Blank matrix: inject the sample background without analyte and confirm the CMD baseline stays flat. A rising CMD on a matrix-only injection signals a ridge–groove mismatch (see Verifying backfill quality). The MD channel is expected to rise as the surface fouls — that is normal.
- Spike recovery: spike a known amount of analyte into the matrix and into buffer; comparing the responses quantifies matrix effects on binding and detection.
- Reference molograms: keep non-binding reference molograms in the layout for an additional, in-matrix subtraction channel.
5. Calibrate in the matrix
For concentration determination, build the standard curve in the same matrix as the unknowns. The specific signal is preserved in complex media, but the calibration slope can drop modestly relative to buffer because of transport and competition — so a matrix-matched standard curve gives accurate back-calculated concentrations. See Concentration Prediction.
6. Regenerate and clean
Complex samples foul the surface even when the coherent signal stays clean, so cleaning is not optional. Regenerate between cycles as needed (Regeneration Conditions) and run the cleaning protocol after complex-sample runs (Maintenance & Troubleshooting). For quantifying the smallest detectable concentration in matrix, see Limit of detection and sensitivity.
Recommended Regeneration Conditions
Regeneration in focal molography refers to the process of removing bound analytes from the mologram surface to allow for chip reuse or sequential measurements. The regeneration strategy must be tailored to the ligand–analyte interaction and the chemical robustness of the surface. The following regeneration solutions have been tested and work for regenerating [oligo|PEG] or [oligo|oligo] surfaces:
- 50 mM NaOH
- 3 M Guanidinium Chloride, 125 mM NaOH
For the standard regeneration timing and follow-up rinse, see the Regeneration row in Recommended Flow Parameters. Multiple regeneration cycles can be applied when incomplete surface regeneration is observed. When sticky proteins cannot be regenerated from the surfaces, incubate with 1× trypsin/EDTA followed by the regeneration solutions above (full-loop injection of 500 µL trypsin at 10 µL/min for 50 min).
Choosing Regeneration Conditions by Bond Type
| Dominant interaction | Disruption strategy | Example conditions |
|---|---|---|
| Electrostatic / ionic | pH shift or high ionic strength | 10 mM Glycine-HCl pH 2.0, or 1 M NaCl |
| Hydrogen bonds | Chaotropic agents or low pH | 3 M GdnCl, 8 M Urea, or 50 mM NaOH |
| Hydrophobic | Mild surfactants | 0.5% SDS, 0.05% Tween 20 at elevated concentration |
| Mixed / unknown | Combination (cocktail) approach | 3 M GdnCl + 125 mM NaOH |
Escalation Strategy
Always start with the mildest effective condition and escalate only as needed. Harsh regeneration risks denaturing the ligand or damaging the antifouling coating on the chip surface:
- Mild: Low pH buffer (Glycine-HCl pH 2.5) or moderate salt (0.5 M NaCl). Try this first for most antibody–antigen interactions.
- Moderate: 50 mM NaOH or 3 M GdnCl. Effective for most protein–protein interactions.
- Strong: Combined 3 M GdnCl + 125 mM NaOH, or enzymatic digestion with trypsin/EDTA for very sticky analytes.
Verify regeneration efficiency by comparing the baseline level before and after treatment. Recovery of > 95% indicates complete regeneration. Recovery < 80% suggests the conditions are too mild or the surface is accumulating irreversibly bound material.
DDI Chips: Regeneration via DNA Dehybridization
DDI chips offer a distinct advantage: the entire ligand layer can be removed by dehybridizing the DNA anchor strands (e.g., with 50 mM NaOH), and a fresh ligand can be re-immobilized for the next cycle. This eliminates the need to develop analyte-specific regeneration conditions and is the recommended approach for MCK experiments on DDI chips.
Regeneration Scouting Protocol
When developing regeneration conditions for click chemistry chips or novel analyte–ligand pairs:
- Immobilize ligand and perform a single analyte injection.
- Inject the mildest regeneration condition.
- Check baseline return — if > 95%, the condition is sufficient.
- If baseline return is poor, repeat with a stronger condition.
- After finding an effective condition, verify that repeated regeneration cycles (5–10×) do not reduce ligand activity (i.e., the binding capacity should remain stable across cycles).
RmaxEstimation & Ligand Density Optimization
Predicting the expected maximum response helps assess whether an experiment is performing as intended and guides ligand density choices.
Expected Rmax Calculation
The theoretical maximum response for a 1:1 interaction can be estimated from:
Rmax,expected = (MWanalyte / MWligand) × immobilization level × valency
where the immobilization level is the capture level observed during ligand immobilization (in pg/mm²), MW values are molecular weights, and valency is the number of binding sites per analyte molecule (typically 1 for Fab fragments, 2 for full IgG antibodies).
For example, if 20 pg/mm² of a 50 kDa ligand is immobilized and the analyte is a 150 kDa monovalent molecule, the expected Rmaxwould be approximately (150/50) × 20 = 60 pg/mm².
Active Fraction Assessment
If the observed Rmax from fitting is significantly lower than the calculated Rmax,expected, it suggests that not all immobilized ligand is active. Common causes include:
- Ligand denaturation during immobilization (especially relevant for click chemistry chips where covalent coupling conditions may affect protein folding).
- Unfavorable orientation — binding sites may be blocked or sterically hindered on the surface.
- Steric crowding at high ligand densities, where neighboring ligand molecules physically prevent analyte access.
An active fraction of 50–80% is typical for well-optimized assays. Values below 30% warrant optimization of immobilization conditions.
Ligand Density vs. Mass Transport Trade-Off
Higher ligand density produces stronger signals but increases the risk of mass transport limitation. Use the per-assay targets below to balance signal strength against kinetic fidelity:
- For kinetic measurements (SCK, MCK): Aim for an Rmax of 10–50 pg/mm². Lower densities ensure that the binding curve reflects intrinsic kinetics rather than diffusion-limited delivery.
- For LOD experiments: Maximize ligand density to achieve the highest signal-to-noise ratio.
- For screening / qualitative assays: Moderate density (50–100 pg/mm² Rmax) balances signal strength and data quality.
On DDI chips, different ligand densities can be tested on the same chip by diluting the DNA-conjugated ligand with unfunctionalized oligonucleotides in different multiplexing groups (e.g., 1:1 ratio for ~50% density, 1:3 for ~25%).
Recommended Pickup Volume
The pickup volume determines how much sample is aspirated from the vial into the autosampler loop before injection. Choosing the right pickup volume is important because the sample plug travels through tubing before reaching the sensor, and diffusion at the leading and trailing edges of the plug (tail diffusion) can dilute the sample and reduce the effective concentration at the sensor surface.
- Kinetic Measurements (SCK, MCK, etc.): An extra pickup volume of 30 µL is recommended. This ensures that the sample concentration remains constant throughout the association phase and that kinetic parameters are not distorted by diffusion artifacts. This extra pickup volume is also preset for the kinetic templates in the software, but can be adjusted based on the expected kinetics and flow rates.
- Immobilizations:A pickup volume of 30 µL can be used to compensate for tail diffusion effects during immobilization. This is particularly relevant at low flow rates (< 20 µL/min) where the sample spends more time in the tubing and diffusion has a larger impact on the concentration profile.
- When to increase: For longer contact times or higher viscosity samples, consider increasing the pickup volume to maintain a uniform concentration at the sensor. Refer to the Inject and Incubate section for guidance on how pickup volume interacts with incubation duration.
The normalized overlay below shows that with an appropriate pickup volume, the diffusion effects from the tubing are effectively compensated. When properly compensated, the association curves from different conditions align and dissociation only starts upon buffer injection, confirming that the analyte concentration at the sensor is constant.
